Oxygen Concentrator Folee (Model: Y007F-10) 10 Liter
৳ 55,000.00
Specifications:
* Recommended maximum oxygen flow: 10L/min. Dual flow can be adjusted individually.
* The flow range when the nominal pressure at the outlet is 7kPa: 0.5L/min~10L/min.
* Apply 7kPa back pressure at the recommended maximum flow rate, and the flow rate reaches the accuracy of ±10% of the indicated flow rate.
* Oxygen concentration when the nominal pressure of the output terminal is zero (the required oxygen concentration level is reached within 30 minutes of the initial start-up): the oxygen production is 0.5L/min~10L/min, and the oxygen concentration is ≥90%.
* Output pressure: 0.02MPa~0.07MPa
* Compressor safety valve release pressure: 280kPa±50kPa
* Noise: ≤ 60dB(A)
* Maximum nebulization rate: ≥ 0. 1mL/min (model with nebulizer)
* Power supply voltage: 220V±22V/110V±11V, 50Hz/60Hz±1Hz
* Input power: 800VA
* Minimum working time: no less than 30 minutes.
* Oxygen outlet temperature: ≤46℃
* Recommendation: The length of the oxygen tube should not exceed 6 meters, and the oxygen tube should not be bent or deflated.
* Net weight: 31kg
* Size: 38×35×60(cm)
Altitude:
* Rising from sea level to 1828m, the oxygen concentration will not decrease, but the oxygen concentration will be lower than 90% When from 1828m to 4000m
Security system:
* Shutdown due to current overload or loose connection
* The n:compressor shuts down at high temperature
* Low oxygen concentration alarm
Electrical classification:
* Class I equipment, Class B application part, operation mode: continuous operation.
Normal working environment:
* Ambient temperature range: 5℃~35℃;
* Relative humidity range: ≤80%;
* Atmospheric pressure range: 860hPa to 1060hPa.
Storage and transportation environment:
* Ambient temperature range: -40℃~+70℃;
* Relative humidity range: 10% ~ 100%, no condensation;
* Atmospheric pressure range: 860hPa to 1060hPa.
packing list:
* Oxygen generator
* Disposable nasal oxygen tube 2 pieces
* Air filter 6 pieces
* 1 fuse
* Operating instructions 1 copy
* Remote control 1
Note: When the storage and transportation temperature is lower than 5 ℃, the oxygen generator should be placed under normal working temperature for more than 4 hours before use.
Reviews